Purpose: for bacteriological and other investigations
Application: in water laboratories, in laboratories of water utility companies, beer and alcoholic beverage plants, canneries, as well as in food industry enterprises, including child nutrition, in medical institutions, state sanitary control and certification centers, laboratories, State Veterinary Inspectorate, veterinary institutions
Refrigerated thermostats provide:
- temperature in the chamber from 5 oC regardless of the ambient temperature;
- uniform distribution of the temperature in the working chamber due to forced air circulation;
- setting and regulation of the temperature from 5 oC to 60 oC;
- Automatic control of the temperature in the working chamber;
- Digital display of the current and set temperature;
- Maintaining the set temperature in the working chamber during short-term power cuts;
- visual control of the process due to the glass door.
Thermostats with cooling have:
- minimum time to reach operating mode;
- Low power consumption;
- optimal weight characteristics;
- Audible and visual alarm of temperature exceeding;
- ease of operation;
- high reliability in operation.
- Parameter Name
Parameter value
Note
1
2
3
1. Maximum temperature range of thermostatting, °C
from +5 to +60
2. Maximum deviation of the temperature of any point of the working volume, located at least 50 mm from the chamber walls, from the set temperature at steady-state thermal mode, °С, not more:
±1,5
Prior to reaching a steady state thermal mode, the temperature in the working chamber may differ from the preset temperature by a larger value.
3. Time of establishing the operation mode at the maximum temperature in the working chamber, min, max
180
4. Time of continuous operation of the thermostat in the automatic mode, h, no less than
500
5. Power consumption of the thermostat, W, not more
250
6. Dimensions of the working chamber, mm, no less:
- length (depth to door)
- breadth
- height
395
390
490
7. Overall dimensions of the thermostat, mm, max:
- length
- breadth
- height
625
530
935
8. Weight of the thermostat, kg, max:
57
Thermostats maintain the temperature in the chamber from 5°C to 60°C regardless of the ambient temperature
TSO-1/80 SPU - volume of the working chamber, l. 80 Casing made of structural steel with polymer coating and stainless steel
TSO-200 SPU - volume of the working chamber, l. 200. Case design - structural steel with polymer coating and stainless steel