HI 8424 portable pH/ORP meter/thermometer (pH/ORP/T)

22.12.2021
2.9K
HI 8424 Handheld instrument with remote (replaceable) electrode for measuring pH, ORP and temperature in water and solutions to monitor water quality in water treatment processes, environmental studies, swimming pools, etc. With automatic pH calibration and thermocompensation.
 
Calibration is automatic on one or two points with three buffer values (pH 4.01, pH 7.01 and pH 10.01). Once the instrument has been calibrated, the buffer values used during the calibration process are displayed on the LCD screen.
 
Specifications
 
Range
pH -2.00 to 16.00
ORP ±699.9 mV; ±1999 mV
Temperature -20.0 to 120.0°C
Resolution
pH 0.01 pH
ORP 0.1 mV; 1 mV
temperature 0.1°C
Error
pH ±0.01 pH
ORP ±0.2 mV; ±1 mV
temperature ±0.4°C
Ph calibration is automatic, 1 or 2 points (4.01, 7.01, 10.01)
Temperature compensation automatic, -20 to 120°C, or manual without temperature sensor
pH offset ±1 pH
pH electrode HI 1230B
Temperature HI 7662-T stainless steel temperature sensor
Input impedance 10^12 Ohm
Battery 9V
Battery life approximately 150 hours of continuous use; automatically turns off after 20 minutes of non-use
Usage Condition
temperature 0 : 50°C
humidity up to 100%
Dimensions 164 x 76 x 45mm
Weight 180 g
Supplied with HI1230B pH electrode, HI7662 temperature probe, HI70004 pH 4.01 buffer solution sachet, HI70007 pH 7.01 buffer solution sachet, HI700601 (2) electrode cleaning solution sachet, battery, protective case and instructions.